Assisted ventilation
MECHANICAL VENTILATION
Mechanical ventilators work
by generating a pressure gradient within the airways opening
the lungs.
A simple model is one of a circuit of continuous gas flow.
If this flow is interrupted the pressure rises in the circuit.
This is transmitted to the lungs and the increase in pressure
results in lung expansion. When the interruption is removed
the pressure in the circuit drops and the baby is allowed
to breath out.
The cycle of flow interruption
is called intermittent, positive pressure ventilation
(IPPV).
The circuit can be designed
to either deliver a set pressure or a set tidal volume. Currently
pressure controlled ventilators are the most frequently used
but tidal volume control is becoming increasingly popular.
During inspiration air
flows rapidly into the lungs. The time constant is the time
taken for 63% of the delivered air to flow – it takes
three time constants to deliver 95% and five to reach 99%
of equilibrium.
On ventilators the inspiratory and expiratory times can be
set. A very short inspiratory time can lead to inadequate
tidal volume whilst too short an expiratory time leads to
gas trapping.
Damage to the larynx and
trachea rarely occurs during intubation. Ventilation can also
injure the larynx, trachea and lungs if the need is prolonged
or a high pressure is required.
Features of
ventilators
- Gas blender
- Allows adjustment of the inspired oxygen concentration
between 21% and 100%.
- Time adjustment
- Allows different inspiratory and expiratory times.
- Pressure gauge
- Measures applied airway pressure.
- Expiratory relief valve
- A safety device which limits the peak inspiratory pressure.
- Humidifier -
Saturates inspired gases with water and delivers it to the
lungs at 37°C.
- PEEP - Sets level of
PEEP – used to maintain functional residual capacity
(FRC).
- Alarms - Warns
of inadvertent disconnections, pressure loss, and incorrect
cycle timing.
Aims of ventilation
CO2
elimination
CO2 elimination depends largely
on the volume of gas that passes in and out of the alveoli
– proportional to minute volume.
CO2 elimination can be increased
by either an increase in tidal volume or frequency.
Tidal volume can be increased by increasing peak inspiratory
pressure or decreasing PEEP. Tidal volume can also be influenced
by inspiratory time and flow rate of the ventilator. Hypercapnia
(too much CO2) can result from
hypoventilation. Therefore adjustments to ventilation can
result in dramatic improvements.
Oxygenation
Oxygenation (when ventilated) is dependent on:
- Oxygen concentration
- Mean airway pressure (average
pressure delivered through the respiratory cycle)
- Peak Inspiratory Pressure (PIP)
is one of the determinants of MAP both these therefore should
perhaps be mentioned. Other factors of MAP are inspiration
time, frequency and PEEP.
Mean airway pressure adjustments
can optimise functional residual capacity (air left in the
lungs at end expiration) – this is a key element in
gaseous exchange throughout the respiratory cycle.
Typical settings
for conventional ventilation
- Oxygen 40-50%
- Inspiratory time 0.3-0.5 secs
- Expiratory time 0.3-0.5 secs
- I:E ratio 1:1
- Rate 40-90/minute. Here large
variations exist. Few centers use 90 with conventional ventilators
and some start out with 30
- Inspiratory pressure (PIP) 16-20cm
H2O
- Positive End Expiratory pressure
(PEEP) 4-5 cm H2O
- Gas flow 5-8 litres/minute
top


|